21. What is different between information (EN60950-1) and medical (EN60601-1) safety standard?
Add time:2017-11-22 Click:436
What is different between information (EN60950-1) and medical (EN60601-1) safety standard?
![](http://www.kreco.com.cn/inc/uploads/ckeditor/Answer.png)
According to safety standard, the leakage current in EN60950-1 Class I cannot exceed 3.5mA; in EN60601-1 cannot exceed 0.3mA.
Others criteria like safe distance and numbers of fuse are also different.